Condec UPC5110 User Manual

Browse online or download User Manual for Equipment Condec UPC5110. Condec UPC5110 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - UPC5100/UPC5110

63257 UPC5100/UPC5110 Portable & Rack-mountable PneumaticPressure Calibration ConsoleCONDEC Sales Phone No. (888) 295-8475 CONDEC Web Site: www.

Page 2 - Contents

Operation 7 Figure 2-3. Pressure or Vacuum Measurement Sequence NOTE: UPC5100 shown, AC Input (7) and Fill Port (4) are on back side of UPC5110 Rack

Page 3

8 UPC5100/UPC5110 Operation & Maintenance Manual 7. With the vacuum pump still running, close the VENT valve (8) and check for system leaks. If t

Page 4 - 1.0 Introduction

Calibration 9 3.0 Calibration Follow the procedure on the following pages for calibrating the UPC5100/UPC5110. NOTES: • When calibrating, the compu

Page 5

10 UPC5100/UPC5110 Operation & Maintenance Manual3.3 Zero/Span CalibrationSelecting the ZERO/SPAN position on the Condec Calibration Module (PN 60

Page 6 - Introduction

Calibration 11NOTES:1. If readings are not stable or are not within ±20% of zero, the zero correction can’t be entered.2. If readings are not stable o

Page 7 - 2.0 Operation

12 UPC5100/UPC5110 Operation & Maintenance ManualWhen Step 11 is reached, the display changes so that the left most status symbol is a lower circl

Page 8 - 2.3 Initial Setup Procedure

Calibration 132. Press the ENTER button on the module. The display reads 0.00.3. Set the current generator for 20 mA output. Press the ENTER button on

Page 9

14 UPC5100/UPC5110 Operation & Maintenance Manual4.0 Maintenance and ServiceThis section outlines the mechanical and basic electrical repair proce

Page 10

Maintenance and Service 15UPC5100 Installation:1. Lift the panel and chassis by first grasping the fill port fitting and test port.2. Gently place panel/

Page 11 - 2.6 Battery Operation

16 UPC5100/UPC5110 Operation & Maintenance Manual4.2.3 ORION-2D Manifold Removal (PN 55286)Tools required: Phillips screwdriver11/32" wrench

Page 12 - 3.0 Calibration

Copyright © 2002 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America. Specifications subject to change without no

Page 13 - 3.3 Zero/Span Calibration

Maintenance and Service 177. Remove the valve stem seats (8) and valve needle seats (9) using the needle-nose pliers.8. Remove the inner and outer O-r

Page 14

18 UPC5100/UPC5110 Operation & Maintenance Manual4. Apply a small amount of fluorinated Krytox grease to the shaft threads and install the shaft (1

Page 15

Maintenance and Service 194.2.8 ORION-2D Manifold - Panel InstallationTools required: 7/16" open end wrenchPhillips screwdriverhex wrench (.061&q

Page 16 - 3.8 Normal Mode Test

20 UPC5100/UPC5110 Operation & Maintenance Manualset screws (23) with the .061" hex wrench.6. Close the COARSE valve by turning the knob inse

Page 17 - 4.0 Maintenance and Service

Maintenance and Service 214.2.10 Chassis Mounted Regulator (PN 55502) - Removal, Installation and Adjustment Tools required: channel locks adjustable

Page 18 - Maintenance and Service 15

22 UPC5100/UPC5110 Operation & Maintenance Manual4. Remove end of nylon tubing that goes to the SUPPLY SELECT assembly, other end stays connected

Page 19

Maintenance and Service 234.2.12 Supply Pressure (PN 59730) and Pressure Limit Monitor (PN 59706) Gauges - Removal and InstallationTools required: 7/1

Page 20 - Maintenance and Service 17

24 UPC5100/UPC5110 Operation & Maintenance Manual4. Fill the cylinder to approximately 1000 PSIG and check all fittings for leaks. If there are no

Page 21

Maintenance and Service 25NOTE: Some vacuum generators are mounted with screws and nuts or adhesive.5. Clean the vacuum generator by using compressed

Page 22 - - Signal TB1-6 (Red wire)

26 UPC5100/UPC5110 Operation & Maintenance Manual4.2.16 Vacuum Control Regulater (PN 59765) - Removal, Installation and AdjustmentThe vacuum contr

Page 23

5.0 Model Number System ... 366.0 Availa

Page 24

Maintenance and Service 2711. Install panel/chassis assembly in its enclosure as described in Section 4.2.1 on page 144.2.17 Inlet Check Valve, Nitrog

Page 25

28 UPC5100/UPC5110 Operation & Maintenance Manual4.2.19 AC Power/EMI Line Filter (PN 58870) - Removal and InstallationTools required: Phillips scr

Page 26 - Maintenance and Service 23

Maintenance and Service 294.2.21 Power Switch (PN 58878) for Non-Battery Units - Removal and InstallationTools required: Phillips screwdriver11/16&quo

Page 27

30 UPC5100/UPC5110 Operation & Maintenance Manual Installation:1. Connect and solder the wires onto their respective switch terminals (Table 4-4).

Page 28

Maintenance and Service 314.2.24 Power Supply Assembly (Battery Units Only) - Removal and Installation120 VAC input (PN 58727); 220 VAC input (PN 5873

Page 29

32 UPC5100/UPC5110 Operation & Maintenance Manual5. Unplug the power cord from the power source.6. Disconnect the voltmeter and reconnect the batt

Page 30 - See Figure 4-4 on page 35

Maintenance and Service 334.3 Orion-2D Valve Assembly Component Parts ListRef #r PN Description Quantity1 57482 Nut, Valve Needle Housing 9/16-18 22 5

Page 31

34 UPC5100/UPC5110 Operation & Maintenance Manual Figure 4-1. ORION-2D, Exploded View1532201242241OR14OR424122913271725/2X25/2X23/2X34524/3X33342

Page 32

35Figure 4-2. Range Select Switch WiringFigure 4-3. Display Select Switch WiringFigure 4-4. N2 Inlet Check Valve Assembly (PN 60263)

Page 33

36 UPC5100/UPC5110 Operation & Maintenance Manual5.0 Model Number System UPC 510

Page 34 - Maintenance and Service 31

Introduction 1 About This Manual This manual is intended for use by service techniciansresponsible for installing and servicing theUPC5100/UPC5110 pn

Page 35

Available Ranges, Conversions and Resolutions 376.0 Available Ranges, Conversions and Resolutions NOTE: Non-standard ranges are available by special o

Page 36

38 UPC5100/UPC5110 Operation & Maintenance Manual7.0 Options - Replacement KitsThere are numerous replacement PN’s mentioned through out manual th

Page 37

Specifications 398.0 SpecificationsPressure Specifications:Pressure ranges: +- 0-28 in Hg+- 0-387.00 in H2O + 0-2771.0 in H2O +- 0-96.00 kPa+ 0-100.00

Page 38 - Figure 4-4. N

40 UPC5100/UPC5110 Operation & Maintenance ManualCarrying Case UPC5100 only:Type: Aluminum case with latched cover and handleMaterial Thickness: 0

Page 39 - 5.0 Model Number System

Maintenance 41 Figure 8-1. CPU & POWER SUPPLY ASSEMBLY, Sheet 1(For Non-Battery Units Only)6/15/00AFAG10/4/00AH12/12/01NOTES:1- FOR SCHEMATIC SE

Page 40

42 UPC5100/UPC5110 Operation & Maintenance Manual Figure 8-2. CPU & POWER SUPPLY ASSEMBLY, Sheet 2 (For Non-Battery Units Only)

Page 41

Maintenance 43 Figure 8-3. POWER SUPPLY ASSEMBLY (For Battery Units Only)

Page 42 - 8.0 Specifications

44 UPC5100/UPC5110 Operation & Maintenance Manual Figure 8-4. CPU ASSEMBLY, Sheet 1 (For Battery Units Only)MATERIAL ORPART ORDWGNOMENCLATURE O

Page 43 - Physical Specifications:

Maintenance 45 Figure 8-5. CPU Assembly, sheet 2 (For Battery Units Only)

Page 44 - Maintenance

46 UPC5100/UPC5110 Operation & Maintenance Manual Figure 8-6. CPU & POWER SUPPLY SCHEMATIC, Sheet 1 (For Non-Battery Units Only)

Page 45

2 UPC5100/UPC5110 Operation & Maintenance Manual • System Calibration: The instruments can be completely calibrated without being removed from th

Page 46

Maintenance 47 Figure 8-7. CPU & POWER SUPPLY SCHEMATIC, Sheet 2 (For Non-Battery Units Only)

Page 47 - CPU AND BATTERY POWER

48 UPC5100/UPC5110 Operation & Maintenance Manual Figure 8-8. POWER SUPPLY SCHEMATIC, (For Battery Units Only)

Page 48

Maintenance 49 Figure 8-9. CPU SCHEMATIC, Sheet 1 (For Battery Units Only)

Page 49

50 UPC5100/UPC5110 Operation & Maintenance Manual Figure 8-10. CPU SCHEMATIC, Sheet 2 (For Battery Units Only)

Page 50

UPC5100/UPC5110 Warranty and Return Policy 51UPC5100/UPC5110 Warranty and Return PolicyIf possible, please save original packing material which is spe

Page 51

52 UPC5100/5110 Operation & Maintenance ManualUPC5100/UPC5110 Return Material Authorization FormThe repair lab is also equipped to do calibrations

Page 52

Introduction 3 The following schematic provides an overview of the UPC5100/UPC5110’s function. Figure 1-1. UPC5100/UPC5110 Flow Diagram

Page 53

4 UPC5100/UPC5110 Operation & Maintenance Manual 2.0 Operation The following sections explain the various procedures for operating the UPC5100/UP

Page 54

Operation 5 2.2 Shop Air Operation 1. Turn the SUPPLY SELECT valve (7) to the SHOP AIR position and connect a shop air hose to the malefill port (

Page 55

6 UPC5100/UPC5110 Operation & Maintenance Manual The display scaling for these current measurements are as follows: NOTE: UPC5100/UPC5110 reads a