Condec UPC5210 User Manual

Browse online or download User Manual for Equipment Condec UPC5210. Condec UPC5210 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - UPC5200 & UPC5210

63259 UPC5200 & UPC5210 Portable & Rack-mountable PneumaticHigh Pressure Calibration Console Operation and Maintenance ManualCONDEC Sales Pho

Page 2 - Contents

Operation 7 2.2. Using the REGULATOR (1), adjust the maximum intensifier pump input pressure, as read by theREGULATED PRESSURE gauge (6), to 1/10 of

Page 3

8 UPC5200/UPC5210 Operation and Maintenance Manual Figure 2-3. Pressure Measurement Sequence (Absolute Only) NOTE: UPC5200 shown, AC Input (7) and In

Page 4 - 1.0 Introduction

Calibration 9 3.0 Calibration Follow the procedure on the following pages for calibrating the UPC5200/UPC5210. NOTES: • When calibrating, the compu

Page 5

10 UPC5200/UPC5210 Operation and Maintenance Manual 3.3 Zero/Span Calibration Selecting the ZERO/SPAN position on the Condec Calibration Module (PN 6

Page 6 - Introduction

Calibration 11 Figure 3-3. Linearity and Hysteresis Calibration Starting with the instrument's lowest pressure range, sequentially perform the t

Page 7 - 2.0 Operation

12 UPC5200/UPC5210 Operation and Maintenance Manual3.5 Shunt Resistor CalibrationInstall the Condec Calibration Module (PN 60109) and select the SHUNT

Page 8

Calibration 133.8 Normal Mode TestAfter completing the above calibration procedures, you must perform a normal mode test. A current generatorcapable o

Page 9

14 UPC5200/UPC5210 Operation and Maintenance Manual4.0 Maintenance & ServiceThis section outlines the mechanical and basic electrical repair proce

Page 10

Maintenance & Service 15Figure 4-1. UPC5200/UPC5210 Wiring Diagram

Page 11

16 UPC5200/UPC5210 Operation and Maintenance Manual4.2.1 Panel/Chassis Removal and InstallationUPC5200 RemovalTools required: Phillips screwdriver1.

Page 12 - 3.0 Calibration

Copyright © 2002 Rice Lake Weighing Systems. All rights reserved. Printed in the United States of America. Specifications subject to change without no

Page 13 - 3.3 Zero/Span Calibration

Maintenance & Service 174. Loosen and remove the six screws that secure the two clamps that hold the intensifier and accumulator.5. Remove the accu

Page 14

18 UPC5200/UPC5210 Operation and Maintenance Manual11. Blow out any chips from the seat area using compressed air.12. While holding the 4-40 tap stead

Page 15 - 3.7 Permanent Data Storage

Maintenance & Service 197. Install the end bushing (12) and tighten until snug using the isolation valve needle housing socket (PN68508) and socke

Page 16 - 3.9 Self-Check Test

20 UPC5200/UPC5210 Operation and Maintenance Manual4.2.7 Accumulator, Intensifier and ORION-3A Manifold, Panel InstallationNOTE: See Figure 2-1 on page

Page 17 - 4.0 Maintenance & Service

Maintenance & Service 212. Energize the unit and let warm up. Turn RANGE SELECT switch to highest range. To adjust theCOARSE valve, go to step 3.

Page 18 - Maintenance & Service 15

22 UPC5200/UPC5210 Operation and Maintenance Manual4. Remove O-ring (PN 58051) and back-up ring (PN 59735).5. Remove accumulator body from vise and pl

Page 19

Maintenance & Service 23Reassembly of Check Valves:NOTES: • Clean all parts with solvent and use shop air hose to remove any dust particles from a

Page 20 - Maintenance & Service 17

24 UPC5200/UPC5210 Operation and Maintenance ManualReassembly of Rings and Seals:NOTE: Clean all parts with solvent and use shop air hose to remove an

Page 21

Maintenance & Service 254.2.11 Regulator (Standard Pneumatic) and Solenoid Removal Solenoid 120 VAC input - PN 56851, Solenoid 220 VAC input - PN

Page 22 - Maintenance & Service 19

26 UPC5200/UPC5210 Operation and Maintenance Manual4.2.13 Regulator (Tescom) and Solenoid RemovalSolenoid 120 VAC input - PN 56851, Solenoid 220 VAC i

Page 23

5.0 Model Number System ... 446.0 Ranges

Page 24 - Maintenance & Service 21

Maintenance & Service 27Turn the PUMP CONTROL switch (20) off when 1,000 PSI has been achieved. Check all fittings forleaks. If there are no leaks

Page 25

28 UPC5200/UPC5210 Operation and Maintenance Manual4.2.16 Test Port Quick-Connect Fitting (PN 59004) and Filter (PN 54188), Removal and InstallationEv

Page 26

Maintenance & Service 29Installation:1. Grease inside cavity on bottom flat (none allowed in hole) of quick-connect fitting (PN 59034). 2. Grease O-

Page 27

30 UPC5200/UPC5210 Operation and Maintenance Manual6. Inspect and clean parts in solvent (de-natured alcohol) and blow-dry with compressed air. Replac

Page 28 - Maintenance & Service 25

Maintenance & Service 31 Connect wires to the three wire side of new line filter as follows: Green/Yellow wire to A

Page 29

32 UPC5200/UPC5210 Operation and Maintenance Manual Terminal ColorNormally open Black(C) co

Page 30 - Maintenance & Service 27

Maintenance & Service 33Installation:1. Connect and solder the wires onto their respective switch terminals: Zero:

Page 31

34 UPC5200/UPC5210 Operation and Maintenance Manual4.3 ORION-3A Valve Assembly (55287) Parts ListThe following table lists the component parts of the

Page 32 - Maintenance & Service 29

Maintenance & Service 35Figure 4-2. ORION-3A Part No. 55287 Exploded View

Page 33

36 UPC5200/UPC5210 Operation and Maintenance Manual4.4 Intensifier Assembly Parts ListTable 4-3. Intensifier Parts ListRef NumberPN Description Quantity

Page 34

Introduction 1 About This Manual This manual is intended for use by service techniciansresponsible for installing and servicingUPC5200/UPC5210 pressu

Page 35

Maintenance & Service 37Figure 4-3. Intensifier Exploded ViewFigure 4-4. Tescom and Standard Pneumatic Regulator Mounting

Page 36

38 UPC5200/UPC5210 Operation and Maintenance Manual Figure 4-5. CPU and Power Supply Assembly, Sheet 1(For Non-Battery Units Only)6/15/00AFAG10/4/00

Page 37

Maintenance and Service 39 Figure 4-6. CPU and Power Supply Assembly, Sheet 1 (For Non-Battery Units Only)

Page 38 - Maintenance & Service 35

40 UPC5200/UPC5210 Operation and Maintenance Manual Figure 4-7. Power Supply Assembly (For Battery Units Only)

Page 39

Maintenance and Service 41 Figure 4-8. CPU Schematic, Sheet 2 (For Battery Units Only)

Page 40 - Maintenance & Service 37

42 UPC5200/UPC5210 Operation and Maintenance Manual Figure 4-9. Pump Control Board Assembly

Page 41 - CPU AND POWER SUPPLY

Maintenance and Service 43 Figure 4-10. Pump Control Board Schematic

Page 42 - Maintenance and Service

44 UPC5200/UPC5210 Operation and Maintenance Manual5.0 Model Number System U P C 5 2

Page 43

Ranges and Resolutions 456.0 Ranges and Resolutions NOTE: Display resolution, 0.02% of selected range, unless it is not devisable by 1, 2, or 5.RANGE

Page 44

46 UPC5200/UPC5210 Operation and Maintenance Manual7.0 Options, Replacement KitsThere are numerous replacement part numbers mentioned throughout manua

Page 45

2 UPC5200/UPC5210 Operation and Maintenance Manual •Calibration Integrity: Tamper-proof design. Once calibrated, numerous safeguards guarantee the in

Page 46

Specifications 478.0 SpecificationsPressure Specifications:Pressure range: Three independent pressure ranges per instrument. See Section 5.0 on page 44

Page 47 - 5.0 Model Number System

48 UPC5200/UPC5210 Operation and Maintenance ManualControl Panel:Material: Aluminum (5052-H32)Thickness: 0.125 in.UPC5200 Finish: Gray enamel paint wi

Page 48 - 6.0 Ranges and Resolutions

UPC5200/UPC5210 Warranty and Return Policy 49UPC5200/UPC5210 Warranty and Return PolicyIf possible, please save original packing material which is spe

Page 49 - 7.0 Options, Replacement Kits

50 UPC5200/UPC5210 Operation and Maintenance ManualUPC5200/UPC5210 Return Material Authorization FormThe repair lab is also equipped to do calibration

Page 50 - 8.0 Specifications

Introduction 3 Figure 1-1 provides an overview of the UPC5200/UPC5210’s function. Figure 1-1. UPC5200/UPC5210 Flow Diagram

Page 51

4 UPC5200/UPC5210 Operation and Maintenance Manual 2.0 Operation 2.1 Pressure Cylinder (PN 59533) Filling Procedure NOTE: Condec strongly recommends

Page 52

Operation 5 Figure 2-1. Initial Setup Procedure NOTE: UPC5200 shown, AC Input (7) and Input Port (4) are on back side of UPC5210 Rack Mountable Calib

Page 53 - WEB SITE: www.4condec.com

6 UPC5200/UPC5210 Operation and Maintenance Manual 4. Using the REGULATOR (1), adjust the maximum intensifier pump input pressure, as read by theREGUL

Related models: UPC5200

Comments to this Manuals

No comments