CONDEC 3030Reference ManualPN 106375
2.1.1.3 Accessing Other MenusUse the MORE option on the Main Menu to access the other menufunctions.Figure 4 - Menu Map8
2.2 Using the BacklightThe backlight is controlled by the dedicated backlight key. It toggleson and off when the key is pressed; this is one of the f
A note on naming convention:If a 'sub-main' menu has subordinate menus, it will henceforth bereferred to as { function} Main Menu. E.g. th
If SAVE or RECALL is selected use the arrow keys to select the set-up location. Then use the save option to store the current set-up intothe selected
Use the DONE option to save the changes and return home. Whena display is deactivated its configuration is retained. When thedisplay is activated it
display will indicate "OL" when an inappropriate pressure is applied.If "OL" is observed on any pressure display, the pressure sho
interface and select [EXT] (external sensor). Since the interfacebetween the calibrator and the module is digital all the accuracy anddisplay resolut
5. Measuring VoltageTo measure voltage use the input terminals in the front of thecalibrator. Select the Volts function on one of the displays. Thec
Figure 97. Performing a Pressure Switch TestFigure 10To perform a switch test, follow these steps:1. Change the setup to Setup 4 (default switch test
Important NOTE: The pressure Switch Test can be performedwith the following functions[P1] ST, [P2] ST, or EXT ST.2. Connect the calibrator to the sw
CONDEC 3030Reference Manual1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.1 Conta
7. At the top of the display it will now read, "SW OPENED AT" andgive you the pressure that the switch opened at. 8. Press the "NEXT&
8.2 Calibrating a Pressure-to-Current TransmitterTo calibrate a pressure-to-current transmitter (P/I), perform thefollowing steps:1. Connect the calib
8.3 Percent Error FunctionThe calibrator features a unique function which can calculatepressure vs. milliamp error as a percentage of the 4 to 20 mA l
4. LOOP POWER can be toggled on/off, select NEXT when done.5. Use SELECT to toggle through the UNIT options, and select NEXTto move on.6. Use the ↑ a
7. Again, use the arrows to set 0% point and select DONE SET whenfinished and the %ERROR mode will be ready to use.Note: The 0% and 100% point will be
To reset the min/max registers simply press the clear key. Theseregisters are also cleared at power-up or when the configuration ischanged. 10. Factor
Setup 4: The lower display is set to [P1] switch test, the otherdisplays are off.Setup 5: The upper display is set to [P1], the middle display is set
12. Remote Operation12.1 Remote InterfaceThe calibrator can be remotely controlled using a PC terminal, or bya computer program running the calibrato
The LEM232 cable is used for RS-232 communications from thecalibrator to a computer. If the computer only has USB type ports, aUSB to RS-232 converter
12.4 Using Commands12.4.1 Command typesRefer to the Section 10.5 on Remote Commands for all availablecommands. The calibrator may be controlled using
1. IntroductionThe CONDEC 3030 is designed to be a simple to use yet veryversatile pressure calibrator. Its two internal pressure sensors, P1 andP2 po
12.4.3 Response Data TypesThe data returned by the calibrator can be divided into four types:IntegerFor most computers and controllers they are decima
12.5 Remote Commands and Error CodesThe following tables list all commands, and their descriptions, thatare accepted by the calibrator. Table 5: Commo
LOCKOUT Locks out the keypad of the calibrator in remote operation LO_ERR Sets the 0% of span limit for percent error mode LO_ERR Returns the 0%
INHG Pressure in inches of mercury at 0°C MMHG Pressure in millimeters of mercury at 0°C KG/CM2 Pressure in kilograms per square-centimeter MMH2
12.6 Entering CommandsCommands for the calibrator may be entered in upper or lower case.There is at least one space required between the command andpa
For example:CPRT_COEFB -5.774999E-07 enters -5.774999E-07 ascoefficient B.CPRT_COEFB?Returns the number, which was entered for the secondcoefficient o
DAMPTurns the dampening function on or off.For example:If you send DAMP ON this will turn the dampening functionon. DAMP?Returns the current state of
ERROR _LOOP?Returns the current state of loop power in percent error mode.For example:If you send ERROR_LOOP? It will return ON if loop poweris on in
FUNC?Returns the current mode of all displays. For example if thecalibrator is set to [P2] ST on the upper display, [P1] on themiddle, and RTD on the
OHMS?Returns the raw Ohm value from the RTD. For example:If when measuring a P100-385 at 0 degrees cel sendingOHMS? would return 1.000000E+02, OHM .PR
Symbol DescriptionAC (Alternating Current)AC-DCBatteryCE Complies with European Union DirectivesDCDouble InsulatedElectric ShockFusePE GroundHot Surfa
ST_CLOSE?Returns the pressure that the switch closed at in the currentpressure units.ST_OPEN?Returns the pressure that the switch opened at in the cur
13. Specifications (18 °C to 28 °C unless otherwise noted.)GeneralInstrument Setup Recall 5; last used on power-upEnvironmentalOperating Temperature
AccuracyPressure0.3 psi ±0.1% F.S.1.0 psi, 5 psi ±0.05% F.S.15 psi through 3000 psi ±0.025% F.S.5000 psi, 10,000 psi ±0.035% F.S.mA ±0.015% of rdg
CautionTo avoid damaging the plastic lens and case, do not use solvents orabrasive cleansers.Clean the calibrator with a soft cloth dampened with wate
Condec 3030 Ranges and ResolutionsRange (PSI) 0.3 1 5 15 30 50 100 300 500 1000 1500 3000 5000 10000Burst Pressure (PSI) 10 10 100 500 500 1000 1000 2
Rice Lake Weighing Systems - CONDEC Products3 Simm Lane Door D, Unit 2A Newtown, CT 06470Tel: (888) 295-8475Fax: (203) 364-1556 or (715) 234-6967 we
WarningTo avoid possible electric shock or personal injury:• Do not apply more than the rated voltage. See specifications forsupported ranges. • Fol
CautionTo avoid possible damage to calibrator or to equipment under test:• Use the proper jacks, function, and range for your measurement orsourcing a
Figure 2 shows the location of the keys. Table 2 describes thefunction of each key.Figure 2 - KeypadTable 2 Key FunctionsNo. Name Description 1 Fun
2.1 Calibrator DisplayThe Calibrator Display consists of two regions: The menu bar(located along the bottom of the screen) is used to access a menusys
Here the SELECT option will toggle through the choices for eachparameter. The first parameter is MODE. Since voltage, current andswitch test modes a