Condec 3030 User Manual

Browse online or download User Manual for Equipment Condec 3030. Condec 3030 User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - CONDEC 3030

CONDEC 3030Reference ManualPN 106375

Page 2

2.1.1.3 Accessing Other MenusUse the MORE option on the Main Menu to access the other menufunctions.Figure 4 - Menu Map8

Page 3 - 1. Introduction

2.2 Using the BacklightThe backlight is controlled by the dedicated backlight key. It toggleson and off when the key is pressed; this is one of the f

Page 4 - Symbol Description

A note on naming convention:If a 'sub-main' menu has subordinate menus, it will henceforth bereferred to as { function} Main Menu. E.g. th

Page 5

If SAVE or RECALL is selected use the arrow keys to select the set-up location. Then use the save option to store the current set-up intothe selected

Page 6 - 2. Calibrator Interface

Use the DONE option to save the changes and return home. Whena display is deactivated its configuration is retained. When thedisplay is activated it

Page 7

display will indicate "OL" when an inappropriate pressure is applied.If "OL" is observed on any pressure display, the pressure sho

Page 8 - 2.1 Calibrator Display

interface and select [EXT] (external sensor). Since the interfacebetween the calibrator and the module is digital all the accuracy anddisplay resolut

Page 9

5. Measuring VoltageTo measure voltage use the input terminals in the front of thecalibrator. Select the Volts function on one of the displays. Thec

Page 10 - Figure 4 - Menu Map

Figure 97. Performing a Pressure Switch TestFigure 10To perform a switch test, follow these steps:1. Change the setup to Setup 4 (default switch test

Page 11 - 2.3 Using the Zero Function

Important NOTE: The pressure Switch Test can be performedwith the following functions[P1] ST, [P2] ST, or EXT ST.2. Connect the calibrator to the sw

Page 12

CONDEC 3030Reference Manual1. Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11.1 Conta

Page 13

7. At the top of the display it will now read, "SW OPENED AT" andgive you the pressure that the switch opened at. 8. Press the "NEXT&

Page 14 - 3. Measuring Pressure

8.2 Calibrating a Pressure-to-Current TransmitterTo calibrate a pressure-to-current transmitter (P/I), perform thefollowing steps:1. Connect the calib

Page 15 - 3.1 Media Compatibility

8.3 Percent Error FunctionThe calibrator features a unique function which can calculatepressure vs. milliamp error as a percentage of the 4 to 20 mA l

Page 16 - 4. Measuring Current

4. LOOP POWER can be toggled on/off, select NEXT when done.5. Use SELECT to toggle through the UNIT options, and select NEXTto move on.6. Use the ↑ a

Page 17 - 5. Measuring Voltage

7. Again, use the arrows to set 0% point and select DONE SET whenfinished and the %ERROR mode will be ready to use.Note: The 0% and 100% point will be

Page 18

To reset the min/max registers simply press the clear key. Theseregisters are also cleared at power-up or when the configuration ischanged. 10. Factor

Page 19

Setup 4: The lower display is set to [P1] switch test, the otherdisplays are off.Setup 5: The upper display is set to [P1], the middle display is set

Page 20 - 8. Calibrating Transmitters

12. Remote Operation12.1 Remote InterfaceThe calibrator can be remotely controlled using a PC terminal, or bya computer program running the calibrato

Page 21

The LEM232 cable is used for RS-232 communications from thecalibrator to a computer. If the computer only has USB type ports, aUSB to RS-232 converter

Page 22 - 8.3 Percent Error Function

12.4 Using Commands12.4.1 Command typesRefer to the Section 10.5 on Remote Commands for all availablecommands. The calibrator may be controlled using

Page 23

1. IntroductionThe CONDEC 3030 is designed to be a simple to use yet veryversatile pressure calibrator. Its two internal pressure sensors, P1 andP2 po

Page 24

12.4.3 Response Data TypesThe data returned by the calibrator can be divided into four types:IntegerFor most computers and controllers they are decima

Page 25 - 10. Factory Setups

12.5 Remote Commands and Error CodesThe following tables list all commands, and their descriptions, thatare accepted by the calibrator. Table 5: Commo

Page 26

LOCKOUT Locks out the keypad of the calibrator in remote operation LO_ERR Sets the 0% of span limit for percent error mode LO_ERR Returns the 0%

Page 27 - 12. Remote Operation

INHG Pressure in inches of mercury at 0°C MMHG Pressure in millimeters of mercury at 0°C KG/CM2 Pressure in kilograms per square-centimeter MMH2

Page 28

12.6 Entering CommandsCommands for the calibrator may be entered in upper or lower case.There is at least one space required between the command andpa

Page 29 - 12.4 Using Commands

For example:CPRT_COEFB -5.774999E-07 enters -5.774999E-07 ascoefficient B.CPRT_COEFB?Returns the number, which was entered for the secondcoefficient o

Page 30

DAMPTurns the dampening function on or off.For example:If you send DAMP ON this will turn the dampening functionon. DAMP?Returns the current state of

Page 31

ERROR _LOOP?Returns the current state of loop power in percent error mode.For example:If you send ERROR_LOOP? It will return ON if loop poweris on in

Page 32 - Units Meaning

FUNC?Returns the current mode of all displays. For example if thecalibrator is set to [P2] ST on the upper display, [P1] on themiddle, and RTD on the

Page 33 - Table 8: Error Codes

OHMS?Returns the raw Ohm value from the RTD. For example:If when measuring a P100-385 at 0 degrees cel sendingOHMS? would return 1.000000E+02, OHM .PR

Page 34 - 12.6 Entering Commands

Symbol DescriptionAC (Alternating Current)AC-DCBatteryCE Complies with European Union DirectivesDCDouble InsulatedElectric ShockFusePE GroundHot Surfa

Page 35

ST_CLOSE?Returns the pressure that the switch closed at in the currentpressure units.ST_OPEN?Returns the pressure that the switch opened at in the cur

Page 36

13. Specifications (18 °C to 28 °C unless otherwise noted.)GeneralInstrument Setup Recall 5; last used on power-upEnvironmentalOperating Temperature

Page 37

AccuracyPressure0.3 psi ±0.1% F.S.1.0 psi, 5 psi ±0.05% F.S.15 psi through 3000 psi ±0.025% F.S.5000 psi, 10,000 psi ±0.035% F.S.mA ±0.015% of rdg

Page 38

CautionTo avoid damaging the plastic lens and case, do not use solvents orabrasive cleansers.Clean the calibrator with a soft cloth dampened with wate

Page 39

Condec 3030 Ranges and ResolutionsRange (PSI) 0.3 1 5 15 30 50 100 300 500 1000 1500 3000 5000 10000Burst Pressure (PSI) 10 10 100 500 500 1000 1000 2

Page 41 - 13. Specifications

Rice Lake Weighing Systems - CONDEC Products3 Simm Lane Door D, Unit 2A Newtown, CT 06470Tel: (888) 295-8475Fax: (203) 364-1556 or (715) 234-6967 we

Page 42 - 15. Maintenance

WarningTo avoid possible electric shock or personal injury:• Do not apply more than the rated voltage. See specifications forsupported ranges. • Fol

Page 43

CautionTo avoid possible damage to calibrator or to equipment under test:• Use the proper jacks, function, and range for your measurement orsourcing a

Page 44

Figure 2 shows the location of the keys. Table 2 describes thefunction of each key.Figure 2 - KeypadTable 2 Key FunctionsNo. Name Description 1 Fun

Page 45

2.1 Calibrator DisplayThe Calibrator Display consists of two regions: The menu bar(located along the bottom of the screen) is used to access a menusys

Page 46

Here the SELECT option will toggle through the choices for eachparameter. The first parameter is MODE. Since voltage, current andswitch test modes a